Safety

Caution

Observe all instructions and cautions in the user documentation. Using the product in a manner not specified can damage the product and compromise the built-in safety protection. Return damaged products to ALIARO for repair.

Safety Voltages

Maximum input voltage

100V

Isolation

100V

Channel-to-channel

100V

Channel-to-earth

Short Circuit Current

500mA (protected by current limiting circuit)

Caution

If you are using the AL-2010 with voltages greater than 30 V RMS, 42 V peak, or 60 V DC, the SLSC chassis must be panel mounted in a closed rack to prevent user access to the rear of the device.

These test and measurement circuits are rated for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS.

MAINS is a hazardous live electrical supply system to which equipment is designed to be connected to for the purpose of powering equipment. This product is rated for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

Safety Compliance Standards

This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

  • IEC 61010-1, EN 61010-1

  • Conforms to UL Standard 61010-1 and UL Standard 61010-2-030

  • Certified to CSA Standard C22.2 # 61010-1 and CSA Standard C22.2 # 61010-2-030

Note

For safety certifications, refer to the product label or the Product Certifications and Declarations section.