Analog Input Characteristics¶
The AL-1032 supports a variety of ways to configure a system that can measure an analog voltage from a device under test. The following sections outline what paths are available.
Supported |
6:1 Attenuation |
|
---|---|---|
Module AI |
||
Ch 1..32 |
Yes |
No |
External AI |
||
Ch 1..16 |
Yes |
Yes |
Ch 17..24 [1] |
Yes |
No |
Ch 25..32 |
No |
Not Applicable |
Module AI Characteristics¶
Each channel on the AL-1032 is capable of monitoring voltage internally.
Voltage Range |
-60 V to +60 V |
Voltage step size |
~1.83 mV/div |
Acquisition details |
|
Acquisition Type |
Single-Point |
Timing Type |
On Demand |
Update Rate |
Determined by system configuration and the NI-SLSC driver |
Name |
Resource Type |
DataType |
Description |
Valid Values |
---|---|---|---|---|
AL.AI |
Physical Channel |
Double |
Analog input value in volts |
[-60.0..60.0] |
External AI Characteristics¶
In addition to the internal ADC, the AL-1032 supports connecting external instrumentation to the RTI.
Input to Output Voltage Ratio |
6 V/V |
Without Attenuation¶
See Passthrough for details