Analog Input Characteristics

The AL-1032 supports a variety of ways to configure a system that can measure an analog voltage from a device under test. The following sections outline what paths are available.

Supported

6:1 Attenuation

Module AI

Ch 1..32

Yes

No

External AI

Ch 1..16

Yes

Yes

Ch 17..24 [1]

Yes

No

Ch 25..32

No

Not Applicable

Module AI Characteristics

Each channel on the AL-1032 is capable of monitoring voltage internally.

AI-M
Specifications

Voltage Range

-60 V to +60 V

Voltage step size

~1.83 mV/div

Acquisition details

Acquisition Type

Single-Point

Timing Type

On Demand

Update Rate

Determined by system configuration and the NI-SLSC driver

SLSC Interface Properties

Name

Resource Type

DataType

Description

Valid Values

AL.AI

Physical Channel

Double

Analog input value in volts

[-60.0..60.0]

External AI Characteristics

In addition to the internal ADC, the AL-1032 supports connecting external instrumentation to the RTI.

Input to Output Voltage Ratio

6 V/V

Without Attenuation

See Passthrough for details

With Attenuation

AI-ATT